Rad-97
Chapter 4: Operation
www.masimo.com
64
Masimo
In Vivo for SpO2
In Vivo is currently not available in the U.S.A. and territories relying on FDA market
clearance.
From the
In Vivo
screen, change any of the following options:
Options
Description
Factory Default
Settings
User Configurable
Settings
Enabled
See
In Vivo Adjustment
Overview
on page 61.
Off
On or Off
Offset
Amount
See
In Vivo Adjustment
Overview
on page 61.
0.0 when turned
on
-6.0% to +6.0%, in steps
of 0.1%
PR Settings
From the
PR Setting
s screen, change any of the following options:
PR Alarms
on page 64
Trends
on page 97
About Parameter Information
on page 78
PR Alarms
From the
PR Alarms
screen, change any of the following options:
Options
Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High Limit High Limit is the upper
threshold that triggers an
alarm.
High
140 bpm
35 bpm to 235 bpm,
in steps of 5 bpm
Low Limit Low Limit is the lower
threshold that triggers an
alarm.
High
50 bpm
30 bpm to 230 bpm,
in steps of 5 bpm
Silence
Duration
Sets the amount of time
that the alarm is silenced.
NA
2 minutes
30 sec, 1, 2 or 5
minutes