Rad-97
Chapter 1: Technology Overview
www.masimo.com
27
Masimo
2.
The amount of arterial blood in tissue changes with pulse
(photoplethysmography). Therefore, the amount of light absorbed by the varying
quantities of arterial blood changes as well.
The Rad-97 uses a multi-wavelength sensor to distinguish between oxygenated blood,
deoxygenated blood, blood with carbon monoxide, oxidized blood and blood plasma.
The Rad-97 utilizes a sensor with various light-emitting diodes (LEDs) that pass light through
the site to a diode (detector). Signal data is obtained by passing various visible and infrared
lights (LEDs, 500 to 1400nm) through a capillary bed (for example, a fingertip, a hand, a foot)
and measuring changes in light absorption during the blood pulsatile cycle. This information
may be useful to clinicians. The maximum radiant power of the strongest light is rated at ≤ 25
mW. The detector receives the light, converts it into an electronic signal and sends it to the
Rad-97 for calculation.
1.
Light Emitting Diodes (LEDs)
(7 + wavelengths)
2.
Detector