Chapter 9: Alarms and Messages
www.masimo.com
130
Masimo
Pi Delta
From the
Pi Delta
menu screen, change any of the following options:
Options
Description
Factory
Default
Settings
User Configurable
Settings
Set
Baseline
Sets the Perfusion Index (Pi) value to be
used as the baseline.
Off
On or Off
Percent
Change
The change in Pi from the baseline that,
if maintained for the Timeout length,
will trigger audible and visual alarms.
50%
10% to 99%, in steps of
1%
Timeout
The length of time over which the
percent change in Pi is monitored.
None
None or 1, 5, 30
minutes, 1, 4, 8, 12, 24,
36, 48 hours
Messages
The following section lists common Rad-97 messages, potential causes, and next steps.
Message
Potential Causes
Next Steps
(Pulse CO-Ox)
Replace Cable
or
(RAM) Replace
Cable
•
The patient cable is
non-functional or the patient
monitoring time of the cable
has expired.
•
Replace the patient cable.
(Pulse CO-Ox)
Cable Near
Expiration
or
(RAM) Cable
Near Expiration
•
Patient cable has less than 10%
of patient monitoring time
remaining.
•
Replace with new patient
cable.