mV
Range: 0.000
– 250.00 mV
Resolution: 0.01 mV
V
Range: 0.000
– 30.000 V
Resolution: 0.001 V
Switch
Test
Potential free contacts
Trigger level : 24V, 24mA (2V)
Voltage level detection
Trigger level : 0 to 30V in 1V steps
Input impedance : >1MΩ
Current Measurement
LC-12 supports current measurement using either LC-12 as the
loop power supply while at the same time measuring the current or
simply measuring the current while an external power supply is
used.
The following picture displays the connection for Current
Measurement for different mode. And
also different ways of
providing the supply power to the loop.
mA Current Measurement
In this mode LC-12 not providing any supply voltage. For
proper measurement the external device should capable of
providing the voltage supply. If the external device should not
capable, an external Power Supply should be connected in
series.