P a g e
|
51
of
52 User Manual
6.2 Measurement Parameters and Accuracy
Parameter Range
Resolution
Accuracy
mV
0-250.00 mV
0.01 mV
+0.02% of r 2 count
V
0-30.00 VDC
0.001 V
+0.02% of r 2 count
mA
0-24.00 mA
0.001 mA
+0.02% of r 2 count
6.3 Source Accuracy
Parameter
Range
Resolution
Accuracy
mV
0-250.00 mV
0.01 mV
+0.02% of r 2 count
V
0-12.00 VDC 0.001 V
+0.02% of r 2 count
mA
0-24.00 mA
0.001 mA
+0.02% of r 2 count
6.4 Testing Features
Feature
Value
Loop power output
24V DC, + 10% (24mA maximum)
Output
impedance
in
HART compatible mode
250Ω + 20%
Switch Test
Potential free contacts
Trigger level : 24V, 24mA (2V)
Voltage level detection
Trigger level : 0 to 30V in 1V steps
Input impedance : >1MΩ