Ref.: WB-VL-65 DSP LX Premium - OM Ver.1.6 (B2927)
35
Remove the Calibration weight from the Outer plane & fix it on the Inner plane
of the wheel directly opposite as shown below:
Fig. 37
Press
key twice to perform Inner calibration. The added weight will be
displayed in the Inner plane window and the Outer plan window will be zero.
Fig. 38
Once Inner calibration is performed, press
key to save the Weight
calibration values. Saving process will happen with message prompts for few
seconds and system will go back to
WEIGHT CALIBRATION
screen.
During Outer or Inner calibration, if “Calibration weight signal is too
high. Press
Key to proceed, or any other key to exit” error
message appears, check whether the correct calibration weight is
added on the appropriate plane or weight is not added in the Outer
plane or not shifted to Inner plane. If the weight is correct, and still
you got this error, press ENTER key and proceed with the calibration
If any error is indicated after adding weight in any of the Plane, redo
the Weight calibration
Remove the Calibration wheel and go the
POSITION TRACKING
screen. Run
the Empty shaft and zero is displayed in both the planes.
Now Balance a Wheel as explained in Chapter 7.4 and then add a known
weight and ensure the weight added & its positions are displayed in the
respective windows correctly.