31
MAGNAMED TECNOLOGIA MÉDICA S/A
Icon
Description
Battery charged and in operation. Equipment disconnected from inoperant electric network
or electric network.
Battery partially charged and in operation. Equipment disconnected from electric network or
inoperant electric network.
Battery with minimum charge and in operation. Equipment disconnected from electric
network or inoperant electric network. Low battery alarm can be activated.
6.3.7 Audible alarm paused button
This indicator appears during the time that audible alarm is silenced (maximum of 2 minutes).
•
If during this period any other alarm appears, alarm pause is automatically deactivated.
6.3.8 Quick Access function area
This area displays buttons that activate the functions of quick access:
Function
Description
STAND BY
Enables or disables Standby mode. In standby mode, the alarms are
stopped, and the ventilation is paused.
As a matter of safety, to enable/disable the standby mode, it is necessary
to press the corresponding button for 1 second to active it.
MUTE ALARM
Keeps the alarms mute by up to 120 seconds.
•
If during this period a new alarm occurs, the mute will be automatically disable.