MAGNAMED TECNOLOGIA MÉDICA S/A
11.13.1 Specifications of adjustable alarms
Table 22
– Adjustable alarms
Alarm
Adjust
Limit
Default parameters
(1)
Unit
Neonatal
Pediatric
Adult
Maximum Pressure
OFF, 0 to 120
High
30
30
40
cmH
2
O
Low
OFF
OFF
OFF
PEEP
OFF, 0 to 80
High
10
15
20
cmH
2
O
Low
OFF
OFF
OFF
Total Volume
OFF, 0 to 3000
High
50 mL
500 mL
1.0 L
L or mL
Low
OFF
OFF
OFF
Minute Volume
OFF, 0.0 to 99
High
1.0
10
20
L
Low
0.5
2
3.6
Rate
OFF, 0 to 200
High
80
60
60
rpm
Low
5
5
5
FiO
2
OFF, 18 to 100
High
80
80
80
%
Low
OFF
OFF
OFF
Apnea time
OFF, 0 to 60
High
15
15
15
s
Automatic adjust
(2)
OFF, 10, 20 and 30
OFF
%
(1) Whenever the equipment is turned on or there is an exchange of patient type or end the power of the
battery without the ventilator is connected to the power grid, the alarms will assume default values.
(2) Applies only to the basic alarms ventilation (maximum pressure, PEEP, volume, minute volume, rate and
FiO
2
).