28
MAGNAMED TECNOLOGIA MÉDICA S/A
Sequence
Procedure
Mode Adjust
By touching on the button that indicates the active ventilation mode, it is
possible to configure the modes available by patient.
As each mode is selected, the adjustable parameters available for this mode
are shown in the central display area. To exit from this screen, confirm the
mode selection and their settings by pressing the button CONFIRM, or discard
all changes by pressing the button CANCEL.
Adjustment buttons
Adjustment
Procedure
Adjustment of
ventilatory
parameters
For adjustment of ventilatory parameters, touch on the button
corresponding to the parameter to be set, it will be selected (the color
will change), allowing the modification of its value using the spin and
confirm button.
To confirm the set value, press the button again corresponding to the
parameter or press the spin and confirm button (ENTER).
Alarms adjustment
To access the alarm settings screen, tap on the ALARM tab.
For setting the alarm, tap the button corresponding to the alarm to be
set, this will be selected (the color will change), allowing the
modification of its value using the spin and confirm button.
To confirm the set value, press the button again corresponding to the
alarm or press the spin and confirm button (ENTER).
•
If a new value were not confirmed, it will be discarded after 5 or 10 seconds of inactivity,
depending on the ventilator model.