MAGNAMED TECNOLOGIA MÉDICA S/A
•
Before first use, the equipment and its components must be cleaned, disinfected and sterilized
as specified.
•
All parts of MAGNAMED ventilators who have contact with fluids from the patient (ex. breathing
circuit) are potentially contaminated, they are called semi critical and must undergo before
being discarded (at the end of their life cycle) or sending maintenance service or repair, a high
level of disinfection or sterilization process.
•
In the case of disposal: INDICATE as medical waste potentially infected.
•
By sending the ventilator for maintenance or repair services, to observe STRICTLY the
disinfection process.
•
Do not use damaged patient cables.
•
Accessories and removable MAGNAMED ventilator components is damaged or signs of wear
should be replaced, avoiding use.
•
The packaging of non-sterile devices (breathing circuits, expiratory connectors and valves) is
designed to keep these products at the appropriate level of cleanliness to be sterilized before its
use and also to minimize microbial contamination.
•
Cleaning
o
Do not soak or immerse the monitor in any fluid.
o
Use sparingly cleaning solution. The excess can drip inside the monitor and cause
internal damage to components.
•
Do not use abrasives to perform cleaning.
•
Do not use alcohol to clean the plastic parts, except when specified.
•
Do not immerse the ventilator in any liquid.