The distal flow sensor must not be sterilized by steam sterilization.
•
For the internal disinfection of DISTAL flow sensor, do not use tools that may generate
mechanical forces, such as air or water jet, at the risk of damaging the filament.
•
Drying of DISTAL flow sensor should occur naturally in ambient air, so avoid the use of
compressed air or dryer.
•
If there is a possibility of the sensor remain infected, replace it promptly.
Sterilization
•
The components that get in touch with the respiratory gases must be removed for cleaning and
sterilization;
•
Do not use abrasive agents to carry out cleaning;
•
Do not use alcohol to clean the plastic parts;
•
Do not immerse the equipment in any liquid.
8.3.1 Steam sterilization
Perform the set sterilization as the autoclave manufacturer's recommendation.
Table 8
– Autoclavable accessories
Description
Autoclave cycles
Breathing circuit
50
Silicon line
50
Exhalation valve
50
Diaphragm of exhalation valve
50
Important Advices