60
MAGNAMED TECNOLOGIA MÉDICA S/A
Connectivity: connection with
computers for rescue and data printing
(tend and events)
Updating software without having to
open the device
RS-232 connector (EIA RS-232C) Type DB9 female
Pneumatic Specifications
11.5.1 Pneumatic Chart
Figure 6
– Pneumatic chart – FlexiMag
11.5.2 Gas inlet connections
Table 15
– Gas inlet connections
Item
Specification
Connections
As ABNT NBR 11906:2011
03
Proportional O
2
02
Reg Pressure
01
Regulated Valve
Oxygen 40~150 PSI
06
Anti-asphyxia
15
NIF Maneuver
05
Nebulizer TGI
08
Oxygen
Concentration
01
– Pressure regulating valve
02
– Regulated pressure measured (Pressure sensor)
03
– Proportional valve control
04
– Unidiretional valve Scheader
05
– Solenoid for overpressure control
06
– Unidiretional valve (Relief valve)
07
– Solenoid to nebulizer TGI control
08
– Unidiretional valve (Anti-Asphyxia)
09
– Pressure intake
10
– O
2
concentration
11
– Inspiratory flow sensor (internal)
12
– Expiration control - PEEP
13
– Expiratory flow sensor (external)
Pressure
sensor
Reg. Valve
AIR Inlet
Prop. Valve
O
2
concentration
Flow sensor
Flow sensor
Exhalation
valve
Pressure
intake
Anti-asphyxia
Nebulizer TGI
Pressure
sensor
O
2
Inlet
Prop. Valve
200 L/min
Reg. Valve
Sc
h
ra
d
e
r
Va
lv
e
Sc
h
ra
d
e
r
Va
lv
e
Overpressure
valve
Solenoid valve
Three ways
Patient