MAGNAMED TECNOLOGIA MÉDICA S/A
Function
Description
O
2
100%
Keeping the oxygen concentration at 100% for 90 or 120 seconds after
pressing. This feature can be used for procedures of pre and post-
aspiration of secretions in the airways.
MANUAL CYCLE
Manually shoots an inspiratory cycle, as selected ventilatory mode.
PAUSE
INSPIRATORY
Allows maneuvers of suspension of inspiratory time, widely used in the case
of X-ray of the chest.
By pressing and immediately releasing this button, the inspiration will be
extended for a minimum period. Keeping pressed, the expiration will be
extended for up to 30s.
After this period, the value of static compliance will be displayed.
PAUSE
EXPIRATORY
Allows extension expiration time maneuvers (prolonged expiration time).
By pressing and immediately releasing this button, the expiration will be
extended for a minimum period that is adjustable. Keeping pressed, the
expiration will be extended for up to 30s.
After this period, the value of static compliance will be displayed.
P0.1
Pressure drop below the baseline pressure generated by the inspiratory
effort of the patient and measured during the first 100ms from the
beginning of the inspiratory phase.
P0.1 is often used as references for weaning the patient.
After the drive, the ventilator will measure P0.1 for 10 seconds or until the
disengagement by the operator. The result will be shown as one of the
main monitor parameters.
LOCKS
OF KEYBOARD
(LOCK)
Protection system against accidental change. Lock or unlock the touch
screen. When commands are crashed on the display, press the button to
unlock them IMMEDIATELY.
To lock again simply press this button once or wait 2 minutes without
touching the screen.
•
Only will be locked the ventilator’s adjustment functions. The navigation in the equipment, the
visualization of parameters and the switching between ventilation and standby status will be released.
•
Some features may be temporarily disabled, depending on the machine status (when in standby status,
for example).