82
MAGNAMED TECNOLOGIA MÉDICA S/A
11.10.8 DualPAP
DualPAP
–
Dual Continuous Positive Airway Pressure
Description:
In DualPAP, the ventilator works at two
pressure levels adjusted by the operator,
Pr Superior and Pr Inferior.
The change to the lower level of pressure
occurs at the end of Superior T (given
time to the higher level of pressure).
Similarly, the restoration of the higher
pressure level so happens that is
exhausted Low T (time for the level of
the lower pressure).
Consequently, the respiratory rate and
I:E ratio are directly related to this
alternation between levels.
DualPAP allows spontaneous cycles in
both pressure levels and includes
synchronization of possibility with the
inspiratory effort of the patient.
When there are no spontaneous
breathing, DualPAP is similar to pressure
control mode, differing from the latter
because the adjusting time (T Upper and
Lower T), rather than the respiratory
rate.
Set parameters:
•
PR. SUPERIOR
•
T. SUPERIOR
•
PR. INFERIOR
•
T. INFERIOR
•
CONCENTRATION
•
ΔPS (Press
ure Support
–
PEEP)
•
LIMIT PRESSURE
•
CYCLING FLOW (% FLOW)
•
TRIGGER PRESSURE
•
TRIGGER FLOW
•
RISE TIME
•
FLOW (
- NEONATAL);
•
BACKUP
o
Backup VCV
▪
VOLUME
▪
RATE
▪
FLOW
▪
LIMIT PRESSURE
o
Backup PCV
▪
PRESSURE INSP
▪
RATE
▪
TIME INSP
▪
RISE TIME
o
Backup PLV-
NEONATAL
▪
PRESSURE INSP
▪
RATE
▪
TIME INSP
o
Backup Auto
(1)
1
–
Whenever the apnea
set time is reached, the
ventilator triggers a
ventilatory cycle whose
configuration is based on
the current mode
settings.