46
MAGNAMED TECNOLOGIA MÉDICA S/A
8.1.2.1.2
Rinse
a) Always use potable water for rinsing;
b) Thoroughly rinse the external surface of the accessories with potable water.
c) Rinse the internal surface by injecting potable water under pressure at least 5 times.
•
Not to be used for cleaning or disinfecting the phenol (> 5%) ketone, formaldehyde, hypochlorite,
chlorinated hydrocarbons, aromatic hydrocarbons, inorganic acids, and quaternary ammonium
compounds.
•
Never use saline solutions, especially sodium hypochlorite (bleach) and saline, disinfectants,
hydrogen peroxide for cleaning or rinsing the accessories.
8.1.2.1.3
Drying
Drying of the external parts should be done with a clean, soft and dry cloth and the drying of the internal
parts should be done so that the solution drains by gravity.
Disinfection
8.2.1 External Parts
The external part should be disinfected using a clean cloth moistened with alcohol 70 º.
8.2.2 Respiratory circuit and exhalation valve
After cleaning, the items should be disinfected with alcohol 70º. The external part should be disinfected
using a clean cloth moistened with alcohol 70 º and internal part by immersion. Important: Do not soak the
items to be disinfected with alcohol as it can damage the material.
After disinfection, the external parts should be dried with a clean, soft and dry cloth and the internal
parts should be dried so that the solution drains by gravity.
8.2.3 DISTAL flow sensor (Envitec SpiroQuant A+)
1. Disconnect the distal sensor from the exhalation valve and the connector cable and wait 30 minutes.
2. Immerse the sensor in 70% ethanol solution for 1 hour
3. Allow to dry naturally for 30 minutes in the environment before reassembly.