90
MAGNAMED TECNOLOGIA MÉDICA S/A
Parameter
Range
Resolution
Tolerance
(1)
O
2
Concentration (FiO
2
)
12,0 to 99,9 %
0,1
±1% in volume or
2% of read
100 to 110 %
1
Dynamics Complacency
0 to 99,9 mL/cmH
2
O
0,1
1mL/cmH
2
O or
10% of the measured
value
100 to 200 mL/cmH
2
O
1
Static Complacency
0 to 99,9 mL/cmH
2
O
0,1
1mL/cmH
2
O or
10% of the measured
value
100 to 200 mL/cmH
2
O
1
Ti / Ttotal
Calculates(s)
0,1
---
RSBi
–
Rapid shallow breathing
index (IRRS, Tobin index)
Calculated(cycles/min/L)
1
---
WOBi (Imposed Work Of
Breathing)
Calculated (J/min)
0,01
---
WOBi (Imposed Work Of
Breathing)
Calculated (J/L)
0,01
---
P0.1
-99 to -10 cmH
2
O
1
---
-9,9 to 0,0 cmH
2
O
0,1
(1) When two tolerances are indicated, consider the higher value.
(2) For airway resistance greater than 150 cmH2O/L/s, the expired volume will have its tolerance changed to
10%. In this condition, the measured inspiratory volume remains unchanged.
Observations
•
In practice, the units of measure of pressure are equivalent, being able to adopt that 1 mbar =
1 hPa ≈ 1 cmH2O.
Specifications of Alarm and Safety System
•
Valve anti-asphyxia to protect against failures in gas supply.
•
Safety relief valve 100 cmH
2
O as basic standard ventilators in order to avoid overpressure in the
breathing circuit safety.
•
Active overpressure valve that does not detect obstructions, is activated to reduce the pressure
in the breathing circuit.