LeadCare Ultra
®
Blood Lead Testing System User’s Guide
7-4
7.3 Summary of Display Messages
Normally, only five messages are displayed during the test process:
•
”INSERT A SENSOR”
•
“SCAN A BARCODE”
•
“LOAD THE SAMPLE”
•
“ANALYZING THE SAMPLE”
•
“RESULT”
However, other messages may appear if the analyzer detects a condition that prevents normal
operation. Table 7-3 shows the possible display messages.
Table 7-3: UA Error Messages
Error Type/Message
Reason(s)
Fix
System Errors
Analyzer Connection to Computer Workstation
•
The analyzer is turned
OFF when UA is open.
•
The analyzer is
disconnected from
power source when
UA is open
•
The analyzer is not
connected to the
computer workstation
by the USB cable.
•
Confirm the power
switch on the back of
the analyzer is set to
the ON position.
•
Confirm the power
socket/outlet
connection to the
analyzer power
supply is set to the
ON position.
•
Confirm that the
analyzer is connected
to the computer via
the USB cable.
Summary of Contents for LeadCare Ultra
Page 16: ...LeadCare Ultra Blood Lead Testing System User s Guide 2 6 This page intentionally left blank ...
Page 62: ...LeadCare Ultra Blood Lead Testing System User s Guide 6 20 This page intentionally left blank ...
Page 92: ...LeadCare Ultra Blood Lead Testing System User s Guide 11 2 This page intentionally left blank ...
Page 96: ...LeadCare Ultra Blood Lead Testing System User s Guide 13 2 This page intentionally left blank ...
Page 102: ...LeadCare Ultra Blood Lead Testing System User s Guide A 4 This page intentionally left blank ...
Page 103: ...LeadCare Ultra Blood Lead Testing System User s Guide B 1 ...
Page 104: ...LeadCare Ultra Blood Lead Testing System User s Guide B 2 ...
Page 116: ...LeadCare Ultra Blood Lead Testing System User s Guide B 12 This page intentionally left blank ...
Page 118: ...LeadCare Ultra Blood Lead Testing System User s Guide C 2 This page intentionally left blank ...