LeadCare Ultra
®
Blood Lead Testing System User’s Guide
4-3
Figure 4-2: LeadCare Ultra UA Main Screen
NOTES:
•
The first time you turn on the analyzer after installation, the “Calibration
Required” dialog box appears:
Figure 4-3: Calibration Required Screen
•
Once the system has been calibrated, the lot number of the calibration
button is displayed in the upper left corner of the main screen
(Sensor Lot #).
•
If the “Sensor Lot #” displayed does not match the kit lot number you
intend to use, you must recalibrate the analyzer with the calibration
button from the kit lot you intend to use.