LeadCare Ultra
®
Blood Lead Testing System User’s Guide
4-5
4.4 Entering Quality Control Settings
Quality control information can be entered anytime. When calibrating the analyzer with a new button,
the “Quality Control Settings” dialog box opens automatically (Figure 4-7).
If the QC lot has expired, the associated expiration date in the “Quality Control Settings” dialog box will
turn red, flagging you to update the QC information.
NOTE:
At any other time, to access the Quality Control Settings, click
Tools
>
Set
the QC Levels
from the UA.
►
Follow These Steps
1. Using the arrows, select the desired number of QC levels. You have the ability to enter up to
three levels of quality control.
Figure 4-7: New QC Lot Dialog Box
2. Enter the information for the new QC lot in the appropriate box:
a. Control Name: Defaults to “LEVEL 1, LEVEL 2”, etc., for each level. To rename, delete
the default name and enter the new name in the field.
b. Control Lot #: Enter the lot number in the field.
c. Target Range: Enter the
target value
in the first field and the
tolerance
in the second
field. For example, a control with a target value of 7.0 + 3.0 is entered as shown in
Figure 4-8.
Summary of Contents for LeadCare Ultra
Page 16: ...LeadCare Ultra Blood Lead Testing System User s Guide 2 6 This page intentionally left blank ...
Page 62: ...LeadCare Ultra Blood Lead Testing System User s Guide 6 20 This page intentionally left blank ...
Page 92: ...LeadCare Ultra Blood Lead Testing System User s Guide 11 2 This page intentionally left blank ...
Page 96: ...LeadCare Ultra Blood Lead Testing System User s Guide 13 2 This page intentionally left blank ...
Page 102: ...LeadCare Ultra Blood Lead Testing System User s Guide A 4 This page intentionally left blank ...
Page 103: ...LeadCare Ultra Blood Lead Testing System User s Guide B 1 ...
Page 104: ...LeadCare Ultra Blood Lead Testing System User s Guide B 2 ...
Page 116: ...LeadCare Ultra Blood Lead Testing System User s Guide B 12 This page intentionally left blank ...
Page 118: ...LeadCare Ultra Blood Lead Testing System User s Guide C 2 This page intentionally left blank ...