LeadCare Ultra
®
Blood Lead Testing System User’s Guide
7-8
Error Type/Message
Reason(s)
Fix
Temperature Below Limit or Temperature Above
Limit During Blood Lead Test
•
The temperature of the
analyzer is outside of
its operational
temperature range.
•
Place the system in
an environment of
ambient temperature
16°C - 30°C. Allow
the analyzer to adjust
to ambient
temperature.
Sensor Out of Vial Too Long
•
Sensor was left in the
analyzer for more than
five (5) minutes
without use. Sensor
performance may be
compromised.
•
Remove and discard
sensor. Insert a fresh
sensor, enter Sample
ID and add sample to
sensor without delay.
Sensor Inserted Before Ready
•
One or more sensors
were detected in the
analyzer as the
analyzer was turned
ON.
•
Remove and re-insert
sensor. If the error
continues, call
Magellan Product
Support.
Sensor(s) Inserted on Exit
•
One or more sensors
were detected in the
analyzer as the
analyzer was turned
OFF.
•
Remove and discard
sensors.
Summary of Contents for LeadCare Ultra
Page 16: ...LeadCare Ultra Blood Lead Testing System User s Guide 2 6 This page intentionally left blank ...
Page 62: ...LeadCare Ultra Blood Lead Testing System User s Guide 6 20 This page intentionally left blank ...
Page 92: ...LeadCare Ultra Blood Lead Testing System User s Guide 11 2 This page intentionally left blank ...
Page 96: ...LeadCare Ultra Blood Lead Testing System User s Guide 13 2 This page intentionally left blank ...
Page 102: ...LeadCare Ultra Blood Lead Testing System User s Guide A 4 This page intentionally left blank ...
Page 103: ...LeadCare Ultra Blood Lead Testing System User s Guide B 1 ...
Page 104: ...LeadCare Ultra Blood Lead Testing System User s Guide B 2 ...
Page 116: ...LeadCare Ultra Blood Lead Testing System User s Guide B 12 This page intentionally left blank ...
Page 118: ...LeadCare Ultra Blood Lead Testing System User s Guide C 2 This page intentionally left blank ...