Excessive Noise in Data
Table 9-3 Troubleshooting Excessive Noise
Cause(s)
Solution(s)
Mechanical vibration
Move or shut down any sources of vibration. Isolate the
ODiSI 6 from those sources.
Dirty or loose fiber
and/or bulkhead
connections
Clean all connections and reconnect the fiber
connections
. For instructions, see “Cleaning Connectors”
in the Maintenance and Cleaning section in the Setup
Guide, which can be accessed from the Help menu on the
ODiSI 6.
Subpar routing and/or
orientation of standoff
cable and/or remote
module
Reroute and reorient the standoff cable and remote
module to minimize signal degradation and to arrive at a
more ideal physical configuration.
Cracks in bulkhead
connector alignment
sleeve
Contact Luna Technologies at 1-866- LUNAOVA or
No Data After Measurement
Table 9-4 Troubleshooting No Data
Cause(s)
Solution(s)
Incorrect graph range
settings
Rescale the plot axes.
Errors on Software Startup
Table 9-5 Troubleshooting Software Setup
Cause(s)
Solution(s)
The USB cable
connections might be
loose
Check the connections to ensure that the cable is
connected properly to both the instrument and the
controller.
The instrument is not
turned on
Ensure that the instrument is plugged into the proper type
of outlet and turn on the power switch.
The software is in an
unknown state
Exit and restart the software.