ODiSI 6
ODiSI 6
User’s Guide
Page 54
Figure 6-36: Filter and Downsample menu item
Spatial Moving Average Size
: Specifies the size of the spatial moving average filter window
to apply to the calculated strain/temperature results.
Spatial Downsampling Factor
: Specifies the number of gage results to throw away spatially,
thereby changing the “effective gage pitch.” For example, providing a value of 3 here means
that for every 3 neighboring gages only 1 of those gage results will be kept.
Effective Gage Pitch
: The gage pitch when taking into account any specified spatial down
sampling. This represents the nominal spacing between reported gage results.
Temporal Moving Average Size
: Specifies the size of the time moving average filter window
to apply to the calculated strain/temperature results.
Temporal Downsampling Factor
: Specifies the reduction factor in time of the reported
strain/temperature results.
Measurement Rate Per Channel
: Measurement rate that will be achieved on each channel.
This rate depends on the sensor length, Measurement Mode, Gage Pitch, Performance Mode,
number of channels, and Temporal Downsampling Factor.
Figure 6-37: Filter and Downsample settings dialog