ODiSI 6
ODiSI 6
User’s Guide
Page 38
Figure 6-9: View measurements down the entire sensor length for a specific channel
Sensor Tare
A sensor can be zero-ed out before the start of a test. This re-zero-ing is called a Tare. Tares
are specific to Gage Pitches.
Save Tare
: Click this button to zero out a sensor.
Tare
dropdown menu
: Select from multiple saved Tares.
Delete Tare
: Click to delete the selected Tare.
Sensor Key
Any sensor whose key has been installed in the software and is able to be identified can be
rekeyed. Reasons for needing to Rekey include re-splicing a broken sensor, taking the sensor
to higher strain ranges, re-centering the strain and temperature range around a new point,
and permanently removing strain gradients induced in the sensor. For example, if a strain
sensor is installed and is loaded to 10,000
µε prior to a test, the system will only be able to
measure an additional 5,000
µε. If the sensor is rekeyed, the sensor can then be used over a
larger strain range. It is important to ensure that the sensor and system are static and in a
quiet environment when Rekeying a sensor.
Rekey
: Click this button to capture a new Rayleigh backscatter reference for this sensor.
Key dropdown menu
: Select from multiple sensor keys to be used for the sensor.
Delete Key
: Click to delete the selected Rekeyed trace.