Performance Mode
: The Performance Mode selection determines the processing capability
of the instrument. Selecting “Maximum Rate” provides the fastest measurement rate given the
Measurement Mode (i.e. remote module type), sensor lengths, and Gage Pitch selection.
Selecting “Full Optimization” provides the highest quality results at half the maximum
measurement rate given the Measurement Mode (i.e. remote module type), sensor lengths,
and Gage Pitch selection.
X-Axis Units
: Select length units from the drop-down menu.
Use Patch Cord
: Check if using a patch cord (fiber jumper) between the remote module and
the sensor. Required to properly identify sensors when using patch cords.
Figure 6-25: General settings dialog
Channel Settings
Select the Channel Setti
ngs menu by clicking “Settings → Channel Settings”.
Figure 6-26: Channel Settings menu item
Standoff Length
: Select standoff cable length per channel from the drop-down menus (10 m,
50 m, 100 m, 150 m, or 200 m). The longer lengths can be achieved by connecting two shorter
standoff cables together using a Luna provided connector.