![Luna ODiSI 6 User Manual Download Page 36](http://html1.mh-extra.com/html/luna/odisi-6/odisi-6_user-manual_1949788036.webp)
ODiSI 6
ODiSI 6
User’s Guide
Page 35
Example: Operational Parameters
Figure 6-6: Sensor Properties screen with three connected sensors
In the above figure:
•
Status:
“Ready”, meaning the hardware has been set up and is ready to identify
sensors and make measurements.
•
Measurement Mode: HD Standard Length selected
•
Gage Pitch: 2.6 mm Gage Pitch selected
•
Performance Mode: Full Optimization selected
•
Measurement rate per channel: Based on the sensor lengths, number of channels,
Gage Pitch, and Performance Mode, the resultant measurement rate per channel is
25.00 Hz
•
Channel 1: A 10.27 m strain sensor has been identified and added to the configuration
of Channel 1 and this channel is ready for making measurements. The Status
message is in green and t
he “View” button is active.
•
Channel 2: A sensor has been identified but not added to the configuration of Channel
2. It is a 20.19 m strain sensor on an Extended Length remote module. This does not
match the Measurement Mode selection of Standard Length. This description is
displayed in the Status of Channel 2 in black.
•
Channel 3: A 5.24 m strain sensor has been identified as being connected to the
remote module through a 3.18 m patch cord. This does not match the original