4.3.1 Basic Menu
Menu item
Description
Adjustable value / option
Ac-
cess
min.
max.
stan-
Unit
value
value
dard
Ready
Operating display
-
-
-
-
-
Freq. Setpoint
Sets the speed for operation
150
*
*
Hz
r/w on
!! Every change is directly written
into the pump’s data storage
and is valid immediately!!
Motor Temp.
Motor temperature
actual value
°
C
r
Converter Temp. Temperature of the power
actual value
°
C
r
electronic
Bearing Temp.
Temperature of the magnetic
actual value
°
C
r
bearing
Actual Value TMS Temperature of the Temperature
actual value
°
C
r
Management System
T_A
Value for Cooling Temp. Control
actual value
°
C
r
Cooling Temp.
Cooling water temperature
actual value
°
C
r
Actual PVW 13
Rotor displacement in the
actual value
%
r
magnetic bearing plane VW13
Actual PVW 24
Rotor displacement in the
actual value
%
r
magnetic bearing plane VW24
Actual PZ 12
Rotor displacement in the
actual value
%
r
magnetic bearing axis Z12
Power
Power consumption of the drive
actual value
W
r
Analog Input 1
Analog Input Channel 1 (0...100.0%)
actual value
%
r
Input range 0...10 V
Displ. 0.00...100.0%
Analog Input 2
Analog Input Channel 2 (0...100.0%)
actual value
%
r
Input range 0...10 V
Displ. 0.00...100.0%
Plug-in control
56
GA 05.141/6.02 - 07/2003
Ready
0.0 A
0 Hz
Enter
Enter
Motor Temp.
50
°
C
Enter
Converter Temp.
40
°
C
Enter
Bearing Temp.
50
°
C
Enter
Actual PVW 13
10%
Enter
Actual PVW 24
10%
Enter
Actual PZ 12
5%
Enter
Power
400 W
Enter
Not Active
Actual Value TMS
70
°
C
Not Active
Cooling Temp.
38
°
C
Enter
r (read)
= value can only be read
r /w (read/write) = value can be read and written
on (online)
= value can be changed always
off (offline)
= value can only be changed when the pump stands still
4.3 Operating menu
Analog Input 1
30.0%
Enter
T_A
68
°
C
Enter
Enter
Analog Input 2
1.5%
Enter
Freq. Setpoint
600 Hz
*
400 Hz = 24,000 rpm for MAG W 830
600 Hz = 36,000 rpm for MAG (W) 1300 & 1500
490 Hz = 29,400 rpm for MAG W 2200
480 Hz = 28,800 rpm for MAG W 2800 & 3200