Installation
24
GA 05.141/6.02 - 07/2003
1
2
1 Aluminum cover
2 Screws
Fig. 15 Removing the transport seal
2.2 Unpacking - storing -
transportation
Remove the equipment from the transportation box and
keep the packaging. Make sure that the product has not
been damaged during transportation. If this unit is dama-
ged contact your carrier and inform LEYBOLD if neces-
sary. For storage of the product, use the packaging pro-
vided.
Lift the pump by the crane eyelets or move it with at least
two persons.
You can position the pump on the base plate for trans-
port with a lift-truck. Protect the pump against slipping
and tipping over.
Caution
Be careful not to damage the sockets and
coolant connections during transportation.
Do not stand below the pump while connecting or remo-
ving the MAG.
The MAG is shipped in a sealed PE bag with desiccant.
Do not open the sealed package until immediately befo-
re installing.
Do not remove the covers and blanking flanges until you
are ready to make the connections, to ensure that the
MAG is installed under the cleanest possible conditions.
2.3 Operating environment
When using the MAG inside a magnetic field, the
magnetic induction at the pump housing must not ex-
ceed 5 mT; (1 mT (milliTesla) = 10 G (Gauß))
Exceeding this limit can cause excessive rotor heating
due to the eddy currents generated in this situation. It is
therefore necessary to provide suitable shielding in such
cases.
The standard version of the MAG is resistant to radiation
at levels up to 10
3
Gy. (1 Gy (Gray) = 100 rad)
The ambient temperature must not exceed 40
°
C
(104
°
F).
The noise level when the pump is running is below
70 dB(A). No acoustic insulation is required.