1.6 Technical data
MAG
W 830 C
W 1300 C
W 1300 C
DN 160
DN 200
DN 250
Pumping speed (PNEUROP)
for N
2
l·s
-1
700
1100
1220
for Ar
l·s
-1
650
1050
1180
for H
2
l·s
-1
300
920
1020
Compression for N
2
> 5·10
7
> 10
8
> 10
8
Ultimate pressure as to DIN 28 400
mbar
< 10
-8
< 10
-8
< 10
-8
Max. forevacuum pressure with N
2
mbar
2
2
2
Rotor Speed
min
-1
24,000
36,000
36,000
Run-up time
min
< 4
< 6
< 6
Braking time with/without venting
min
1 / < 4
1 / < 6
1 / < 6
Cooling
water
Cooling connection, threads
G
1/4“
with Swagelok for tube OD
mm
6.4 (1/4“)
6.4 (1/4“)
–
or with John Guest fitting for tube
mm
–
6
6
Cooling water temperature
°
C
10 to 30
°
F
50 to 86
Weight
approx. kg
32
High-vacuum connection flange
DN
160 ISO-F
200 ISO-F
250 ISO-F
Max. high-vacuum flange temperature
for continuous operation
°
C
85
°
F
185
Bake-out temperature at
high-vacuum flange
°
C
120
°
F
248
Vibration level at high-vacuum flange
at max. speed
µm
< 0.01
Forevacuum connection flange
DN
40 KF
40 KF
40 KF
or
25 KF
Recommended backing pump
Dry commpressing pump with
pumping speed
m
3
/h
100
or rotary vane pump
TRIVAC
D 65 BCS
Admissible ambient temperature
°
C
5 to 40
°
F
40 to 104
Storage temperature
°
C
-10 to +60
°
F
14 to 140
Max. relative air humidity
95% (non-condensing)
Degree of protection (EN 60529)
IP 20
Description
10