Plug-in control
61
GA 05.141/6.02 - 07/2003
4.3.5 Menu Set Pump
Menu item
Description
Adjustable value / option
Ac-
cess
min.
max.
stan-
Unit
value
value
dard
Normal Operation Threshold for normal operation
corresponding to the
35
99
95
%
r/w off
frequency setpoint
Max. Accel. Time Monitor time for acceleration
and overload
600
3600
1000
s
r/w off
4.3.6 Menu Set TMS
Menu item
Description
Adjustable value / option
Ac-
cess
min.
max.
stan-
Unit
value
value
dard
TMS Setpoint
Setpoint of the Temperature-
20
80/90*
70
°
C
r/w on
Management System
4.3.7 Menu Set Purge / Vent
Menu item
Description
Adjustable value / option
Ac-
cess
min.
max.
stan-
Unit
value
value
dard
Purge / Vent
The operator can select one
purge OFF
-
r/w on
of the beside standing
purge ON
-
options for purge and vent
Controlling purge / vent
via control plug X14
-
Option P / V
Displays the factory installed
Installed option
-
r
option for purge and vent
r (read)
= value can only be read
r /w (read/write) = value can be read and written
on (online)
= value can be changed always
off (offline)
= value can only be changed when the pump stands stillstands still
*
MAG (W) 1500: 90
°
C, MAG W 2800, 3200: 80
°
C
Set Pump
Enter
Normal Operation
95 %
Enter
Max. Accel. time
1000 s
Enter
Set TMS
Enter
TMS Setpoint
70
°
C
Enter
Via X14
Purge ON
Set Purge/Vent
Enter
Purge/Vent
Purge OFF
Enter
Option P/V
Option 1
Enter