HistoCore Arcadia H
Operation
5
5.4
Operation modes
Standby mode
If the scheduler is enabled, the instrument enters operation mode and standby mode automatically at
the programmed Start time and End time.
During the standby mode:
• All heating elements (paraffin tank/dispenser, working surface/forceps holder, and trays) are
switched off.
• Cold spot cooling is switched off.
• The dispenser control is disabled.
• Only the
Operate/Standby
button, the
Enhance heating
button, the
Light
button and the
Scheduler
icon (if it is already enabled the day before) are shown on the control panel.
Enabling operating mode
Fig. 21
20
• Touch the
Operate/Standby
button no less than 1 second.
• The instrument switches from standby mode to operating mode. The current time and icons of the
currently selected values are shown on the control panel.
24-hour mode – working in shift mode
If the Start time and End time are set to the same value, the instrument will run continuously, even on
days that are not defined as work days.
Example: Start time =
00:00
and End time =
00:00
.
To set the scheduler,
p. 29 – 5.3 Control panel functions)
Warning
Important
The scheduler is disabled if the end time is before the start time. The scheduler icon is not displayed.
For example, Start time: 08:00 and End time: 06:00.
Note
In 24-hour mode, activate enhance heating mode
p. 34 – 5.5 Instrument heater)
the melting process after replenishing solid paraffin.