Troubleshooting
7
No.
Error
Message
Description
Instrument Behavior
User action
8
2_31
The temperature of the
left tray is higher than the
upper temperature limit.
1. Stop left tray heating
2. Error Message &
Error Icon blinking
3. Alarm beep
Take out the specimen
from the instrument and
contact customer service.
9
2_32
The temperature of the
left tray drops below the
lower temperature limit of
the target.
Error Message &
Error Icon blinking
Contact customer service.
10
2_33
The left tray cannot reach
target temperature after
5 hours the system starts
heating up.
Error Message &
Error Icon blinking
Contact customer service.
11
2_35
The temperature of the
left tray rises too fast.
Error Message &
Error Icon blinking
1. Take out the specimen
from the instrument.
2. Make sure that the
input voltage is right
for the instrument.
3. Contact customer
service.
12
2_41
The temperature of the
right tray is higher than
the upper temperature
limit.
1. Stop right tray heating
2. Error Message &
Error Icon blinking
3. Alarm beep
Take out the specimen
from the instrument and
contact customer service.
13
2_42
The temperature of the
right tray drops below the
lower temperature limit of
the target.
Error Message &
Error Icon blinking
Contact customer service.
14
2_43
The right tray cannot
reach target temperature
after 5 hours the system
is switched on.
Error Message &
Error Icon blinking
1. Make sure that the
input voltage is right
for the instrument.
2. Contact customer
service.
15
2_45
The temperature of the
right tray rises too fast.
Error Message &
Error Icon blinking
1. Take out the specimen
from the instrument.
2. Make sure that the
input voltage is right
for the instrument.
3. Contact customer
service.