28
Version 1.8, Revision K
Operation
5
5.2
Switching the instrument on
Fig. 15
9
20
To switch on the instrument for the first time, follow these steps.
• Fill the paraffin tank with paraffin.
• Set the power switch (right-hand side, next to control panel) to "I" position
.
• The instrument performs a self-test. All icons on the touchscreen light up shortly and go out. Then,
4 buttons, Up, Right, Down, Left, are shown on the screen repeatedly in sequence for a few seconds.
When the 4 buttons go out, the
Operate/Standby
button
button is shown on the screen. The instrument enters standby mode.
• Touch the
Operate/Standby
button no less than 1 second, and the instrument enters operating mode.
Note
• The normal modes of the instrument are Standby and Operating. Switch between the two modes
by using the
Operate/Standby
button.
• Use the power switch only if the instrument is to be switched off for an extended period or to
apply new settings of the Start time and End time.
• Set the temperature of the heating area, the working day, local time, Start time and End time. To set
these values,
p. 29 – 5.3 Control panel functions)
.
• The heaters become active. The melting indicator
flashes once per second during the
heating phase. The temperature settings can be modified during the heating phase.
Note
Prior to leaving the factory, the HistoCore Arcadia H is tested thoroughly under laboratory conditions.
Please check the paraffin tank and the dispenser before using or after service. There will be a small
quantity of clean paraffin in the paraffin tank or dropped from the dispenser. You can work with this
paraffin without any problem.
Warning
• During operation, it is recommended not to fill the instrument with solid paraffin when the paraffin
in the paraffin tank is already melted.
• Risk of burning!
• Risk of blockage in the paraffin tank where it connects with the filler tube.
• Do not add more than 4 L paraffin into the tank.