18
Version 1.8, Revision K
Setting up the instrument
4
4.
Setting up the instrument
4.1
Site requirement
• Stable, vibration-free laboratory table with horizontal, flat table top, as far as possible vibration-free
ground.
• The instrument must not be placed in the vicinity of the air outlet of an air conditioner and must be
shielded from strong sunlight (window).
• To ensure a fully functional heat sink, there must be gap of at least 15 cm behind the instrument.
• The instrument must be installed in a place that ensures an easy disconnection from the power
supply. The power cable must be in a place that can be easily reached.
• The vicinity of the work area must be free of oil and chemical vapors.
Warning
The installation location must be well-ventilated and free of ignition sources of any kind.
The instrument should not be operated in hazardous locations.
4.2
Standard delivery – packing list
Qty.
Designation
Order No.
1
Basic unit HistoCore Arcadia H,
220-240 VAC
14 0393 54090
220-240 VAC, China
14 0393 54103
100-120 VAC
14 0393 54100
220-240 VAC
14 0393 57257
220-240 VAC, China
14 0393 57259
220-240 VAC
14 0393 57258
2
Left/right trays, removable
14 0393 53684
2
Lids for left/right trays
14 0393 53720
1
Paraffin scraper
14 0393 53643
1
Forceps holder, removable
14 0393 55225
1
Paraffin tank filter
14 0393 53559
4
Sets of spare fuses, 250 V 10 A
14 6000 04975
1
Instructions for Use (printed English,
with language CD 14 0393 81200)
14 0393 81001
The country specific power cord needs to be ordered separately. Please find a list of all power cords
available for your device on our website www.LeicaBiosystems.com within the product section.
Note
Please compare the delivered components against the packing list, delivery note, and your order.
Should there be any discrepancy, please contact the Leica Biosystems distributor handling your
order.