30
Version 1.8, Revision K
Operation
5
Note
If no touch action is received within 60 seconds, the setting mode exits automatically. All the settings
are stored until they are changed, even if the instrument is switched off with the power switch.
Setting the temperature
Fig. 18
1
2
3
1
Temperature of trays
3
Temperature of working
surface/forceps holder
2
Temperature of paraffin
tank/dispenser
The temperatures for the instruments different heating areas can be adjusted separately from 50 °C
(122 °F) and 75 °C (167 °F) in 1 °C (or 1 °F) increments.
Warning
When setting the temperature, please observe the paraffin manufacturer's specifications for the
maximum permissible temperature.
1. Touch the
Setup
button and the temperature of the trays will flash.
2. Set the temperature value with the
Up
/
Down
buttons.
Press the
Up
/
Down
button for more than 2 seconds, and the setting value is changed
continuously. Touch the
Left
Right
button to switch between
the trays temperature, paraffin tank/dispenser temperature, working surface/forceps holder
temperature, temperature unit (°C or °F).
3. If necessary, touch the
Up
/
Down
buttons to select Celsius
degree (°C) or Fahrenheit degree (°F).
4. Touch the
Enter
button to save the setting.
After 75 °C (167 °F) has been reached, the temperature will revert to 50 °C (122 °F). Once set, the
temperature value for one range will be retained until it is changed.