24
Version 1.8, Revision K
Operation
5
• The dispenser handle can be operated by simply pushing the mold (or a finger) against the pressure
clip. Slightly pushing the handle backwards will open the dispenser valve
released, the handle springs back in the original position and the valve shuts.
Note
The flow cannot be stopped completely with the metering screw
. It must not be turned
while cold!
Warning
Do not use the dispenser when the instrument is powered off. Otherwise, mechanical damage will be
caused to the dispenser.
Fig. 11
3
4
5
15
4
3
5
Working surface
• The working surface includes the embedding area
and the
.
• The temperature of the embedding area
and forceps holder
can be
adjusted between 50 °C (122 °F) and 75 °C (167 °F) in 1 °C (or 1 °F) increments.
• The working area has grooves and several drain holes
, through which excess paraffin
rapidly drains.
• The cold spot is an integral part of the work area.
• To orientate the specimen, the mold is filled about one third with liquid paraffin. The liquid paraffin
begins to solidify rapidly on the cold spot.
• While the paraffin is semi-liquid, the specimen can be oriented as required. Finally, the mold can
rapidly be filled up with paraffin.