Troubleshooting
7
No.
Error
Message
Description
Instrument Behavior
User action
16
2_51
The temperature of
the working surface is
higher than the upper
temperature limit.
1. Stop working surface
heating
2. Error Message &
Error Icon blinking
3. Alarm beep
Shut down the instrument
and contact customer
service.
17
2_52
The temperature of
the working surface
drops below the lower
temperature limit of the
target.
Error Message &
Error Icon blinking
1. Make sure that the
input voltage is right
for the instrument.
2. Contact customer
service.
18
2_55
The temperature of the
working surface rises too
fast.
Error Message &
Error Icon blinking
1. Stop embedding
workflow.
2. Make sure that the
input voltage is right
for the instrument.
3. Contact customer
service.
19
2_71
The temperature of the
cold spot is below the
lower temperature limit.
Error Message &
Error Icon blinking
1. Make sure that the
room temperatrure is
20~30 °C.
2. Contact customer
service.
20
2_72
The temperature of
the cold spot rises
higher than the upper
temperature limit.
Error Message &
Error Icon blinking
1. Make sure that the
room temperatrure is
20~30 °C.
2. Contact customer
service.
21
2_73
The temperature of the
cold spot is higher than
the upper temperature
limit after 1 hour the
system is switched on.
Error Message &
Error Icon blinking
1. Make sure that the
room temperatrure is
20~30 °C.
2. Contact customer
service.
22
2_61
The temperature of
the forceps holder is
higher than the upper
temperature limit.
1. Stop fordeps holder
heating
2. Error Message &
Error Icon blinking
3. Alarm beep
Take out the forceps from
the forceps holder and
contact customer service.
23
2_62
The temperature of
the forceps holder
drops below the lower
temperature limit of the
target.
Error Message &
Error Icon blinking
Contact customer service.