23
6. Assembly
6.3 Tube and Eyepieces
The tube is mounted to the stand either directly or
with the use of intermediate modules. It is fastened
in place with the side clamping screw (12.1).
• Only for the motorized tube MBDT:
Remove the transportation lock (11.1) from the
bottom side of the tube.
• Loosen the clamping screw (12.1). on the
stand.
• Insert the tube in the circular receptacle
(dovetail ring).
• Retighten the clamping screw (12.1).
• Only for the motorized tube MBDT:
Connect the tube to the stand with the con-
nector socket (13.1).
• The eyepieces are inserted into the eyepiece
tubes on the tube.
Note:
For eyepieces that are not included in shipment,
we recommend to learn them in with the Soft-
ware Leica DMControl, module: Configuration.
This ensures that the information about total
magnification on the LeicaScreen is correct.
Fig. 12
Fastening the tube
1
Clamping screw
1
Fig. 13
Motorized tube connection
1
Connector socket
1
Fig. 11
Bottom side of the tube
1
Transportation lock
1