46
7. Startup
1
Fig. 45
Stage with specimen holder
1
Specimen holder
2
Condenser height adjuster
2
Fig. 44
Remote Control Module SmartMove
1
Move in x-direction
2
Move in y-direction
3
Focus
4
Height adjustment
5
Variable function keys (factory pre-set)
1
2
3
4
5
7.5 The Remote Control Element SmartMove
Turning knobs at the SmartMove
The specimen stage can be moved in x/y
direction with the turning knobs (44.1, 44.2). The
image can be focused with the turning knob
(44.3).
The height of the turning knobs can be adjusted
to your individual needs with the screw (44.4).
Variable function keys at the SmartMove
The function keys are assigned logical functions
before delivery, corresponding to the
configuration of your microscope. The keys are
labeled accordingly. For the setting of the keys,
see ‘Identification Sheet’.
For the list of abbreviations see page
→
41f.
Note:
The setting of the variable function keys can
only be altered via the Software Leica
DMControl, module: Configuration.
7.6 Köhler Illumination
7.6.1 Transmitted Light
For each objective, reasonable values for the
aperture diaphragm and the field diaphragm are
already set. The condenser is also pre-adjusted
in the factory.
However, it may be necessary to re-adjust the
condenser in some cases. Therefore, check the
condenser centering.
The following procedure is provided for the
transmitted light-bright field illumination.