98
Instructions for Use V 2.4 RevL - 12/2015
5. Operation
5.2 Reagents
5.2.1 Editing the reagent list
Entering reagent names
This is done via the
REAGENTS
menu option in the
SETTINGS
menu. You must be logged on at
supervisor access level to proceed.
• From the start screen, press
SETTINGS
to go to the window of the same name and press the
REAGENTS
button there.
• The
SET UP REAGENTS AND WARNING THRESHOLDS
window opens (
Fig. 136
).
• To add a reagent, proceed as follows:
• Press
INSERT
to display the keyboard.
• Enter the new reagent name.
• Press
OK
to confirm.
• You will automatically be prompted to select the reagent group for the new reagent
(
Fig. 137
).
• Allocate the new reagent to the appropriate group and press
OK
to confirm.
Fig
. 136
Select the reagent group
Fig. 137
The
ASP6025
is shipped together with a standard reagent list and predefined infiltration programs.
The reagent list is adapted to these programs. If the operator writes his own programs, new reagents may have
to be entered.