ASP6025 Tissue Processor
4. Instrument Setup
Unpacking the instrument
(
continued
)
Push the instrument from the pallet
(
Fig. 11.3
)
Fig. 11.3
• Hold the rear of the
ASP6025 securely by both
handles (
26
).
(
Fig. 11.2
)
• After the instrument has
been rolled off the pallet,
it can be brought to its final
location.
The left and right illustrations
show how the instrument must
be held when rolled off the pallet
via the ramp.
• When the instrument is set
up in its final location, the
brakes on the instrument
castors must be engaged
once again.
To do so, push the lever
(
97
) (detail in
Fig. 10
) back
downwards.
• When rolling the ASP6025
down from the pallet, sup-
port it by the outer top
corners using both hands
(
Fig. 11.1
).
Fig. 11.1
Fig. 11.2
26
26
Caution!
The instrument castors move very easily. The empty weight of the ASP6025 is 210 kg!
Therefore, it is mandatory to have at least TWO persons hold the instrument when it is rolled off the pallet via
the ramp.