Instructions for Use V 2.4 RevL - 12/2015
6.
Cleaning and Maintenance
Running cleaning programs
(continued)
• After the cleaning program has been completed, a message appears on the
screen indicating that the sensors in the retort have to be cleaned (
Fig. 222
).
Fig. 222
• After you touch the corresponding button, a message appears indicating
that before the cleaning is started the retort must be empty and all speci-
mens have to be removed from it.
• When all necessary conditions have been satisfied, press the
YES
button
to start the cleaning program.
The procedure is displayed graphically on the
SMART SCREEN
.
Fig. 221
Fig. 223
• If threshold values in the reagent status have been exceeded, a warn-
ing message to this effect appears at the end of the cleaning program
(
Fig. 223
).
• If this message is confirmed with
YES
, the
REAGENT STATUS
window
opens, in which you can check the threshold values.
The magnetic stirrer and sieve in the retort must be
removed and thoroughly cleaned regularly after an infiltra-
tion program. Then, the magnetic stirrer and sieve must be
reinserted correctly.