91
ASP6025 Tissue Processor
This function is used to transfer instrument data from a USB stick
into an ASP6025. The following data is saved:
• all programs,
• the reagent list,
• the instrument status,
• the instrument configuration,
• all log files,
• the instrument installation.
Saving or loading data
(continued)
• The instrument software will check that the instrument
data on the stick is actually complete. If it is not, the loading
operation will be terminated.
Fig. 119
Loading data from a USB stick
To write the instrument data from USB stick back to the instru-
ment, proceed as follows:
• Insert a USB stick in one of the USB ports on the instrument
and press the
LOAD
button.
• Press the
ENTRIES
and
LOAD FROM USB
buttons to open
the window of the same name.
• Once all data has been loaded successfully, a message
confirming this appears on the screen and prompts the
user to check that the reagents have been assigned correctly
(
Fig. 120
).
• Finally, successful completion of the entire operation is also
indicated in a message.
Fig. 120
5. Operation
Since there is already instrument data on the
ASP6025, this will be overwritten and thus
deleted. This will be indicated in a warning
message (
Fig. 119
) and must be confirmed
with YES. Pressing the NO button terminates
the process.