ASP6025 Tissue Processor
5. Operation
When connecting the hose, be sure to press it
onto the O-rings of the drain opening as far as
it will go.
The paraffin drain hose must be securely
hooked into the external receptacle and re-
main there through the entire drainage pro-
cess.
When drainage is finished, the hose is
cleaned automatically using air.
Do not remove the hose from the external re-
ceptacle until this cleaning step is complete.
57
• Using the
SMART SCREEN
button, call up the
SMART
FUNCTIONS
screen (
Fig. 181
).
Fig. 180
5.4.3 Draining a paraffin bath
• Connect the paraffin drain hose (
36
) to the spout
(
57
) of the paraffin drain on the front of the instrument
and guide it into a catch tank (
Fig. 180
).
36
Fig. 181
Always drain a paraffin bath using the Re-
mote Drain function only. When doing so, you
should use a suitable vessel with a capacity of
at least 10 liters to drain a 5-liter paraffin bath
to the outside (danger of splashing!).