53
ASP6025 Tissue Processor
4. Instrument Setup
• The drip tray (
16
) is located below the reagent cabinet. It
collects reagents that have overflowed or spilled so that no
pollution can form in or under the instrument. The volume
is approx. 5 l.
• The drip tray (
16
) must be regularly checked for signs of
reagents that have leaked. To do so, pull the tray out by the
handle (
Fig. 53
) and empty it if necessary.
Reagent cabinet with system bottles
(
continued
)
4.4.9 The drip tray
16
Fig. 53
Condensate bottle
• The condensate bottle (
14
,
Fig. 52
) has the exact same shape
and connection design as the system bottle. The difference
is that it is made of blue plastic.
• It captures and collects condensate that forms in the system.
A mark for the maximum level is on the front side.
• Regularly check and empty the condensate bottle (once a
week) (see
Chapter 6.2
.2).
Fig. 52
14
Dispose of waste solvents with care according
to local regulations and the waste manage-
ment policy of the company or institution.