110
Instructions for Use V 2.4 RevL - 12/2015
5. Operation
In the following program (that is to say when another infiltra-
tion program is started in Concentration mode), the ethanol is
replaced.
For this purpose, fresh ethanol must be added manually to RTU-
bottle D4 in the drawer prior to the start of the program.
The operator is prompted to do this by a corresponding screen
instruction (
Fig. 152
).
This must be confirmed with
YES
before the program is started.
The ethanol is then transferred to system bottle S6 while the
program is running.
Functional description of ethanol rotation
(continued)
If
NO
(
Fig. 152
) is pressed, the adjacent message is displayed.
Pressing
YES
(
Fig. 154
) starts the program - WITHOUT ethanol
rotation! - Refilling system bottle S6 is done manually and must
be confirmed in the reagent status (
Fig. 146
, p. 105).
Pressing
NO
stops the program or does not start it.
Fig. 153
Fig. 152
Fig. 154
2nd step
Pressing
OK
(
Fig. 153
) starts the program - fresh ethanol is used.
If
CANCEL
is pressed, the program is stopped or does not start.