User Manual LW120
Parameter range
<freq>
10e3 to 50e6
<ampl>
10e-3 to 10
<offs>
-4.5 to +4.5
<delay>
0 to 99.9
<rise>
0 to 99.9
<fall>
0 to 99.9
APPLy:RAMP?
Response
The LW120 will return <freq>, <ampl>,<offs>,<delay>,<high>,<fall>. Default values are
1e6,5,0,10,10,10
APPLy:SINC{<freq>,<ampl>,<offs>,<N_cycles>}
Purpose
This command is a high-level command that programs the LW120 to output sinc waveform along with its
associated parameters
•
<freq> sets the output frequency in units of hertz
•
<ampl> sets the output amplitude in units of volts
•
<offs> sets the output offset in units of volts
•
<N_cycles> sets the number of zero crossing
Parameter type
Numeric
Parameter range
<freq>
10e3 to 50e6
<ampl>
10e-3 to 10
<offs>
-4.5 to +4.5
<N_cycle> 4 to 100
APPLy:SINC?
Response
The LW120 will return <freq>, <ampl>,<offs>,<N_cycle>. Default values are
1e6,5,0,10
APPLy:EXPonential{<freq>,<ampl>,<offs>,<exp>}
Purpose
This command is a high-level command that programs the LW120 to output exponential waveform along
with its associated parameters
•
<freq> sets the output frequency in units of hertz
•
<ampl> sets the output amplitude in units of volts
•
<offs> sets the output offset in units of volts
•
<exp> sets the exponent value
Parameter type
Numeric
Remote Programming Reference 4-16