User Manual LW120
Response and default
RI
The LW120 will return CONT, T G, or GATE depending on its present setting. Default is CONT.
SWEep:TRIGger:SLOPe(POSitive|NEGative}
Purpose
This command will select slope sensitivity for the sweep trigger input.
•
POSitive will select rising edge
•
NEGative will select falling edge
Parameter type
Discrete
SWEep:TRIGger:SLOPe?
Response and default
E
The LW120 will return POS, or N G depending on its present setting. Default is POS.
SWEep:MARKer<mark_sclk>
Purpose
This command will set the marker sample clock frequency for the sweep mode.
•
<mark_sclk> will set the marker sample clock frequency
Parameter type
Numeric
Parameter range
<mark_sclk>
100e-3 to 100e6 in units of samples per second
SWEep:MARKer?
Response and default
The LW120 will return the present sweep sample clock frequency value. The returned value will be in
standard scientific format (for example: 1KHz would be returned as 1E3 – positive numbers are unsigned).
Default value is 64e6.
VOLTage{<ampl>|MINimum|MAXimum}
Purpose
This command programs the peak to peak amplitude of the output waveform. The amplitude is calibrated
when the source impedance is 50
Ω
.
•
<ampl> sets the amplitude
•
ude to its lowest value
<MINimum> sets the amplit
•
<MAXimum> sets the amplitude to its highest value
Parameter type
Numeric
Parameter range
<ampl>
10e-3 to 10 in units of volts
<MINimu
10e-3 in units of volts
m>
Remote Programming Reference 4-28