User Manual LW120
Connecting multiple instruments in master/slave configuration is
extremely easy. All you need is two types of cables, the first being a
standard BNC to BNC coax cable to feed the sample clock output
and a 9-pin to 9-pin cable to feed the master-to-slave, or slave-to-
slave controls. You can get the standard BNC cable from any store
however, the master/slave cable must be bought from your LeCroy
distributor. Figure 3-27 shows the connection between master and
slave instruments. Additional slave units can be daisy-chained in a
similar way. What you have to remember is that SCLK OUT feeds
an adjacent SCLK IN. Also, always have the last slave unit in the
chain terminated with the special 9-pin terminator.
!
SCLK IN
SCLK OUT
REF IN
RS232
MASTER/SLAVE
FUSE
LINE
FRONT PANEL UTILITY MENU
CONFIGURE GPIB ADDRESS WITH
AM IN
TRIG/FSK IN
SINE OUT
CAUTION
NO OPERATOR SERVICABLE PARTS INSIDE
DO NOT REMOVE COVER
REFER SERVICING TO QUALIFIED PERSONNEL
WARNING
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD
REPLACE FUSE WITH THE SAME TYPE AND RATING
DISCONNECT LINE CORD BEFORE REPLACING FUSE
!
IEEE std 488.2 PORT
AC VOLTAGE
85-265V
47-440Hz
40VA MAX
AUTOMATIC
SELECTION
'T' 0.5A/230V
!
SCLK IN
SCLK OUT
REF IN
RS232
MASTER/SLAVE
FUSE
LINE
FRONT PANEL UTILITY MENU
CONFIGURE GPIB ADDRESS WITH
AM IN
TRIG/FSK IN
SINE OUT
CAUTION
NO OPERATOR SERVICABLE PARTS INSIDE
DO NOT REMOVE COVER
REFER SERVICING TO QUALIFIED PERSONNEL
WARNING
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD
REPLACE FUSE WITH THE SAME TYPE AND RATING
DISCONNECT LINE CORD BEFORE REPLACING FUSE
!
IEEE std 488.2 PORT
AC VOLTAGE
85-265V
47-440Hz
40VA MAX
AUTOMATIC
SELECTION
'T' 0.5A/230V
MASTER
MASTER / SLAVE
LAST SLAVE
TERMINATOR
SLAVE
CABLE
Figure 3-27. Wiring Diagram Master to Slave
Hard wiring the instruments from the back is not enough. You now
have to set up the first instrument in the chain as master and the
other instruments as slaves. Refer to Figure 3-28 and the following
description to learn how to set LW120 instruments as master and
slaves.
1. Connect your instruments as shown in Figure 3-27. Make sure
you designate the first instrument in the chain as master. Also
make sure the last slave is terminated with the 9-pin terminator.
2. On each of the instruments press the UTILITY button
3. Select and press the Couple button on all units
Using The Instrument 3-42