<memory#> will select the memory cell where front panel settings will be recalled from.
Parameter type and range
Integers from 1 through 10
RESet,
*RST
Purpose
This command will reset the LW120 to factory defaults.
TEST?,
*TST?
Purpose
This query will invoke the LW120 selftest function.
Response to query
The LW120 will return “0” for no error, or error code. Error codes are listed below.
Error Code Description
0
No errors, selftest passed
1
CPU and peripherals communication failure
2
Channel 1 failure
4
Channel 2 failure
8
Sequence generator channel 1 failure
16
Sequence generator channel 2 failure
32
Channel 2 sample clock divider failure
*IDN?
Purpose
This query will interrogate the identity of the LW120.
Response to query
The LW120 will return LeCroy,LW120,0,1.0.