ARM:SLOPe{POSitive|NEGative}
Purpose
This command will set the edge sensitivity of which the LW120 will stop in arm mode. The stop and start
commands are applied at the rear-panel TRIG IN connector.
•
POSitive will set the positive edge
•
NEGative will set the negative edge
Parameter type
Discrete
ARM:SLOPe?
Response and default
The LW120 will return POS, or NEG depending on the present setting. Default value is POS.
ARM:BREakpoint:POSition<position>
Purpose
This command will program the breakpoint position for the arm function.
•
<position> will set the breakpoint position
Parameter type
Numeric (integer only)
Parameter range
<position>
0 to 999999 in units of waveform points. The breakpoint position can be programmed in
increments of 4 points minimum.
ARM:BREakpoint:POSition?
Response and default
The LW120 will return the present breakpoint value. Default value is 0.
INITiate:CONTinuous{OFF|ON|0|1}
Purpose
This command will set the output in continuous operation and interrupted operation. The trigger subsystem
commands will affect the LW120 only after it will be set to interrupted operation.
•
ON or 1 will set the interrupted mode
•
OFF or 0 will set the continuous mode
Parameter type
Discrete
INITiate:CONTinuous?
Response to query version
The LW120 will return 1 if the output is interrupted, or 0 if the output is continuous. Default value is 0.
Remote Programming Reference 4-41